19.1.3 Variable capacitance
transducers
Like variable inductance, the
principle of variable capacitance is used in displacement measuring transducers
in various ways. The three most common forms of variable capacitance transducer
are shown in Figure 19.3. In Figure 19.3(a), the capacitor
plates are formed by two concentric,
hollow, metal cylinders. The displacement to be measured is applied to the
inner cylinder, which alters the capacitance. The second form, Figure 19.3(b),
consists of two flat, parallel, metal plates, one of which is fixed and one of
which is movable. Displacements to be measured are applied to the movable plate,
and the capacitance changes as this moves. Both of these first two forms use
air as the dielectric medium between the plates. The final form, Figure
19.3(c), has two flat, parallel, metal plates with a sheet of solid dielectric
material between them. The displacement to be measured causes a capacitance
change by moving the dielectric sheet.
Inaccuracies as low as ±0.01% are
possible with these instruments, with measurement resolutions of 1 micron.
Individual devices can be selected from manufacturers’ ranges that measure
displacements as small as 10-11 m or as large as 1 m. The fact that
such instruments consist only of two simple conducting plates means that it is
possible to fabricate devices that are tolerant to a wide range of
environmental hazards such as extreme temperatures, radiation and corrosive
atmospheres. As there are no contacting moving parts, there is no friction or
wear in operation and the life expectancy quoted is 200 years. The major
problem with variable capacitance transducers is their high impedance. This
makes them very susceptible to noise and means that the length and position of
connecting cables need to be chosen very carefully. In addition, very high
impedance instruments need to be used to measure the value of the capacitance. Because
of these difficulties, use of these devices tends to be limited to those few
applications where the high accuracy and measurement resolution of the
instrument are required.
19.1.4 Variable inductance
transducers
One simple type of variable inductance
transducer was shown earlier in Figure 13.4. This has a typical measurement
range of 0–10 mm. An alternative form of variable inductance transducer shown
in Figure 19.4(a) has a very similar size and physical appearance to the LVDT,
but has a centre-tapped single winding. The two halves of the winding are
connected, as shown in Figure 19.4(b), to form two arms of a bridge circuit
that is excited with an alternating voltage. With the core in the central
position, the output from the bridge is zero. Displacements of the core either
side of the null position cause a net output voltage that is approximately
proportional to the displacement for small movements of the core. Instruments
in this second form are available to cover a wide span of displacement measurements.
At the lower end of this span, instruments with a range of 0–2 mm are
available, whilst at the top end, instruments with a range of 0–5 m can be
obtained.
19.1.5 Strain gauges
The principles of strain gauges were covered earlier in Chapter 13. Because of their very small range of measurement (typically 0–50 µm), strain gauges are normally only used to measure displacements within devices like diaphragm-based pressure sensors rather than as a primary sensor in their own right for direct displacement measurement. However, strain gauges can be used to measure larger displacements if the range of
displacement measurement is extended
by the scheme illustrated in Figure 19.5. In this, the displacement to be
measured is applied to a wedge fixed between two beams carrying strain gauges.
As the wedge is displaced downwards, the beams are forced apart and strained,
causing an output reading on the strain gauges. Using this method,
displacements up to about 50 mm can be measured.
19.1.6 Piezoelectric transducers
The piezoelectric transducer is
effectively a force-measuring device that is used in many instruments measuring
force, or the force-related quantities of pressure and acceleration. It is
included within this discussion of linear displacement transducers because its
mode of operation is to generate an e.m.f. that is proportional to the distance
by which it is compressed. The device is manufactured from a crystal, which can
be either a natural material such as quartz or a synthetic material such as
lithium sulphate. The crystal is mechanically stiff (i.e. a large force is
required to compress it), and consequently piezoelectric transducers can only
be used to measure the displacement of mechanical systems that are stiff enough
themselves to be unaffected by the stiffness of the crystal. When the crystal
is compressed, a charge is generated on the surface that is measured as the
output voltage. As is normal with any induced charge, the charge leaks away
over a period of time. Consequently, the output voltage–time characteristic is
as shown in Figure 19.6. Because of this characteristic, piezoelectric
transducers are not suitable for measuring static or slowly varying
displacements, even though the time constant of the charge–decay process can be
lengthened by adding a shunt capacitor across the device.
As a displacement-measuring device,
the piezoelectric transducer has a very high sensitivity, about one thousand
times better than the strain gauge. Its typical inaccuracy is ±1% of full-scale
reading and its life expectancy is three million reversals.
19.1.7 Nozzle flapper
The nozzle flapper is a displacement
transducer that translates displacements into a pres[1]sure
change. A secondary pressure-measuring device is therefore required within the
instrument. The general form of a nozzle flapper is shown schematically in
Figure 19.7. Fluid at a known supply pressure, Ps, flows through a
fixed restriction and then through
a variable restriction formed by the
gap, x, between the end of the main vessel and the flapper plate. The body
whose displacement is being measured is connected physically to the flapper
plate. The output measurement of the instrument is the pressure Po in the
chamber shown in Figure 19.7, and this is almost proportional to x over a
limited range of movement of the flapper plate. The instrument typically has a
first order response characteristic. Air is very commonly used as the working
fluid and this gives the instrument a time constant of about 0.1 seconds. The
instrument has extremely high sensitivity but its range of measurement is quite
small. A typical measurement range is ±0.05 mm with a measurement resolution of
š0.01 µm. One very common application of nozzle flappers is measuring the
displacements within a load cell, which are typically very small.
No comments:
Post a Comment
Tell your requirements and How this blog helped you.