INTRODUCTION TO
MEASURING INSTRUMENTS
1.1
Definition
of instruments
An instrument is a device in which we can determine the magnitude or value of the quantity to be measured. The measuring quantity can be voltage, current, power and energy etc. Generally instruments are classified in to two categories.
1.2 Absolute
instrument
An absolute instrument determines the magnitude of the quantity to be measured in terms of the instrument parameter. This instrument is really used, because each time the value of the measuring quantities varies. So we have to calculate the magnitude of the measuring quantity, analytically which is time consuming. These types of instruments are suitable for laboratory use. Example: Tangent galvanometer.
1.3 Secondary
instrument
This instrument determines the value of the quantity to be measured directly. Generally these instruments are calibrated by comparing with another standard secondary instrument.
Examples of such instruments are voltmeter, ammeter and wattmeter etc. Practically secondary instruments are suitable for measurement.
1.3.1 Indicating
instrument
This instrument uses a dial and pointer to determine the value of measuring quantity. The pointer indication gives the magnitude of measuring quantity.
1.3.2 Recording
instrument
This type of instruments records the magnitude of the quantity to be measured continuously over a specified period of time.
1.3.3 Integrating
instrument
This type of instrument gives the total amount of the quantity to be measured over a specified period of time.
1.3.4
Electromechanical indicating instrument
For satisfactory operation electromechanical indicating instrument, three forces are necessary. They are
(a) Deflecting force
(b) Controlling force
(c)Damping force
1.4 Deflecting force
When there is no input signal to the instrument, the pointer will be at its zero position. To deflect the pointer from its zero position, a force is necessary which is known as deflecting force. A system which produces the deflecting force is known as a deflecting system. Generally a deflecting system converts an electrical signal to a mechanical force.
1.4.1 Magnitude
effect
When a current passes through the coil (Fig.1.2), it produces a imaginary bar magnet. When a soft-iron piece is brought near this coil it is magnetized. Depending upon the current direction the poles are produced in such a way that there will be a force of attraction between the coil and the soft iron piece. This principle is used in moving iron attraction type instrument.
If two soft iron pieces are place near a current carrying coil there will be a force of repulsion between the two soft iron pieces. This principle is utilized in the moving iron repulsion type instrument.
1.4.2 Force between a
permanent magnet and a current carrying coil
When a current carrying coil is placed under the influence of magnetic field produced by a permanent magnet and a force is produced between them. This principle is utilized in the moving coil type instrument.
1.4.3 Force between
two current carrying coil
When two current carrying coils are placed closer to each other there will be a force of repulsion between them. If one coil is movable and other is fixed, the movable coil will move away from the fixed one. This principle is utilized in electrodynamometer type instrument.
1.5 Controlling force
To make the measurement indicated by the pointer definite (constant) a force is necessary which will be acting in the opposite direction to the deflecting force. This force is known as controlling force. A system which produces this force is known as a controlled system. When the external signal to be measured by the instrument is removed, the pointer should return back to the zero position. This is possibly due to the controlling force and the pointer will be indicating a steady value when the deflecting torque is equal to controlling torque.
Td=Tc (1.1)
1.5.1 Spring control
Two springs are attached on either end of spindle (Fig. 1.5).The spindle is placed in jewelled bearing, so that the frictional force between the pivot and spindle will be minimum. Two springs are provided in opposite direction to compensate the temperature error. The spring is made of phosphorous bronze.
When a current is supply, the pointer deflects due to rotation of the spindle. While spindle is rotate, the spring attached with the spindle will oppose the movements of the pointer. The torque produced by the spring is directly proportional to the pointer deflectionθ .
TC
∝ θ
(1.2)
Td , the pointer will come to=The deflecting torque produced Td proportional to ‘I’. When TC a steady position. Therefore
θ ∝ I
(1.3)
Since, θ and I are directly proportional to the scale of such instrument which uses spring controlled is uniform.
1.6 Damping force
The deflection torque and controlling torque produced by systems are electro mechanical. Due to inertia produced by this system, the pointer oscillates about it final steady position before coming to rest. The time required to take the measurement is more. To damp out the oscillation is quickly, a damping force is necessary. This force is produced by different systems.
(a) Air friction damping
(b) Fluid friction damping
(c) Eddy current damping
1.6.1 Air friction
damping
The piston is mechanically connected to a spindle through the connecting rod (Fig. 1.6). The pointer is fixed to the spindle moves over a calibrated dial. When the pointer oscillates in clockwise direction, the piston goes inside and the cylinder gets compressed. The air pushes the piston upwards and the pointer tends to move in anticlockwise direction.
If the pointer oscillates in anticlockwise direction the piston moves away and the pressure of the air inside cylinder gets reduced. The external pressure is more than that of the internal pressure. Therefore the piston moves down wards. The pointer tends to move in clock wise direction.
1.6.2 Eddy current
damping
1.7 Permanent Magnet
Moving Coil (PMMC) instrument
One of the most accurate type of instrument used for D.C. measurements is PMMC instrument.
Construction: A permanent magnet is used in this type instrument. Aluminum former is provided in the cylindrical in between two poles of the permanent magnet (Fig. 1.7). Coils are wound on the aluminum former which is connected with the spindle. This spindle is supported with jeweled bearing. Two springs are attached on either end of the spindle. The terminals of the moving coils are connected to the spring. Therefore the current flows through spring 1, moving coil and spring 2.
Damping: Eddy current damping is used. This is produced by aluminum former. Control: Spring control is used.
Principle of
operation
When D.C. supply is given to the moving coil, D.C. current flows through it. When the current carrying coil is kept in the magnetic field, it experiences a force. This force produces a torque and the former rotates. The pointer is attached with the spindle. When the former rotates, the pointer moves over the calibrated scale. When the polarity is reversed a torque is produced in the opposite direction. The mechanical stopper does not allow the deflection in the opposite direction. Therefore the polarity should be maintained with PMMC instrument.
If A.C. is supplied, a reversing torque is produced. This cannot produce a continuous deflection. Therefore this instrument cannot be used in A.C.
Torque developed by
PMMC
Let
Td =deflecting torque
TC = controlling torque
θ = angle of deflection
K=spring constant
b=width of the coil
l=height of the coil or length of coil
N=No. of turns
I=current
B=Flux density
A=area of the coil
The force produced in the coil is given by
F= BIL sin θ (1.4)
When θ=90°
For N turns, F=NBIL (1.5)
Torque produced Td = F× ⊥r distance (1.6)
Td = NBIL × b = BINA (1.7)
Td = BANI (1.8)
Td ∝ I (1.9)
Advantages
Torque/weight is high
Power consumption is less
Scale is uniform
Damping is very effective
Since operating field is very strong, the effect of stray field is negligible
Range of instrument can be extended
Disadvantages
Use only for D.C.
Cost is high
Error is produced due to ageing effect of PMMC
Friction and temperature error is present
1.7.1 Extension of
range of PMMC instrument
Case-I: Shunt
A low shunt resistance connected in parrel with the ammeter to extent the range of current. Large current can be measured using low current rated ammeter by using a shunt.
Principle of
operation
When D.C. supply is given to the moving coil, D.C. current flows through it. When the current carrying coil is kept in the magnetic field, it experiences a force. This force produces a torque and the former rotates. The pointer is attached with the spindle. When the former rotates, the pointer moves over the calibrated scale. When the polarity is reversed a torque is produced in the opposite direction. The mechanical stopper does not allow the deflection in the opposite direction. Therefore the polarity should be maintained with PMMC instrument.
If A.C. is supplied, a reversing torque is produced. This cannot produce a continuous deflection. Therefore this instrument cannot be used in A.C.
Torque developed by
PMMC
Let
Td =deflecting torque
TC = controlling torque
θ = angle of deflection
K=spring constant
b=width of the coil
l=height of the coil or length of coil
N=No. of turns
I=current
B=Flux density
A=area of the coil
The force produced in the coil is given by
F= BIL sin θ (1.4)
When θ=90°
For N turns, F=NBIL (1.5)
Torque produced Td = F× ⊥r distance (1.6)
Td = NBIL × b = BINA (1.7)
Td = BANI (1.8)
Td ∝ I (1.9)
Advantages
Torque/weight is high
Power consumption is less
Scale is uniform
Damping is very effective
Since operating field is very strong, the effect of stray field is negligible
Range of instrument can be extended
Disadvantages
Use only for D.C.
Cost is high
Error is produced due to ageing effect of PMMC
Friction and temperature error is present
1.7.1 Extension of
range of PMMC instrument
Case-I: Shunt
A low shunt resistance connected in parrel with the ammeter to extent the range of current. Large current can be measured using low current rated ammeter by using a shunt.
Let Rm =Resistance of meter
Rsh =Resistance of shunt
Im = Current through meter
Ish =current through shunt
I= current to be measure
∴Vm=Vsh
(1.10)
Shunt resistance is made of manganin. This has least thermoelectric emf. The change is resistance, due to change in temperature is negligible.
Case (II): Multiplier
A large resistance is connected in series with voltmeter is called multiplier (Fig. 1.9). A large voltage can be measured using a voltmeter of small rating with a multiplier.
No comments:
Post a Comment
Tell your requirements and How this blog helped you.