2.2 Static characteristics of
instruments :
Accuracy of measurement is thus one
consideration in the choice of instrument for a particular application. Other
parameters such as sensitivity, linearity and the reaction to ambient
temperature changes are further considerations. These attributes are
collectively known as the static characteristics of instruments, and are given
in the data sheet for a particular instrument. It is important to note that the
values quoted for instrument characteristics in such a data sheet only apply
when the instrument is used under specified standard calibration conditions.
Due allowance must be made for variations in the characteristics when the
instrument is used in other conditions.
The various static characteristics
are defined in the following paragraphs.
2.2.1 Accuracy and inaccuracy (measurement uncertainty)
The accuracy of an instrument is a
measure of how close the output reading of the instrument is to the correct
value. In practice, it is more usual to quote the inaccuracy figure rather than
the accuracy figure for an instrument. Inaccuracy is the extent to which a
reading might be wrong, and is often quoted as a percentage of the full-scale
(f.s.) reading of an instrument. If, for example, a pressure gauge of range
0–10 bar has a quoted inaccuracy of š1.0% f.s. (š1% of full-scale reading),
then the maximum error to be expected in any reading is 0.1 bar.
This means that when the instrument
is reading 1.0 bar, the possible error is 10% of this value. For this reason,
it is an important system design rule that instruments are chosen such that
their range is appropriate to the spread of values being measured, in order
that the best possible accuracy is maintained in instrument readings. Thus, if
we were measuring pressures with expected values between 0 and 1 bar, we would
not use an instrument with a range of 0–10 bar. The term measurement
uncertainty is frequently used in place of inaccuracy.
No comments:
Post a Comment
Tell your requirements and How this blog helped you.