2.2.9 Sensitivity to disturbance
All calibrations and specifications of an instrument are only valid under
controlled conditions of temperature, pressure etc. These standard ambient
conditions are usually defined in the instrument specification. As variations
occur in the ambient temperature etc., certain static instrument
characteristics change, and the sensitivity to disturbance is a measure of the
magnitude of this change. Such environmental changes affect instruments in two
main ways, known as zero drift and sensitivity drift. Zero drift is sometimes
known by the alternative term, bias.
Zero drift or bias describes the effect where the zero reading of an
instrument is modified by a change in ambient conditions. This causes a
constant error that exists over the full range of measurement of the
instrument. The mechanical form of bathroom scale is a common example of an
instrument that is prone to bias. It is quite usual to find that there is a
reading of perhaps 1 kg with no one stood on the scale. If someone of known
weight 70 kg were to get on the scale, the reading would be 71 kg, and if
someone of known weight 100 kg were to get on the scale, the reading would be
101 kg. Zero drift is normally removable by calibration. In the case of the
bathroom scale just described, a thumbwheel is usually provided that can be
turned until the reading is zero with the scales unloaded, thus removing the
bias.
Zero drift is also commonly found in instruments like voltmeters that are
affected by ambient temperature changes. Typical units by which such zero drift
is measured are volts/°C. This is often called the zero drift coefficient
related to temperature changes. If the characteristic of an instrument is
sensitive to several environmental parameters, then it will have several zero
drift coefficients, one for each environmental parameter. A typical change in
the output characteristic of a pressure gauge subject to zero drift is shown in
Figure 2.7(a).
Sensitivity drift (also known as scale factor drift) defines the amount
by which an instrument’s sensitivity of measurement varies as ambient
conditions change. It is quantified by sensitivity drift coefficients that
define how much drift there is for a unit change in each environmental
parameter that the instrument characteristics are sensitive to. Many components
within an instrument are affected by environmental fluctuations, such as
temperature changes: for instance, the modulus of elasticity of a spring is
temperature dependent. Figure 2.7(b) shows what effect sensitivity drift can
have on the output characteristic of an instrument. Sensitivity drift is
measured in units of the form (angular degree/bar)/°C. If an instrument suffers
both zero drift and sensitivity drift at the same time, then the typical
modification of the output characteristic is shown in Figure 2.7(c).
Example 2.2
A spring balance is calibrated in an environment at a temperature of 20°C
and has the following deflection/load characteristic
Determine the zero drift and sensitivity drift per °C change in ambient
temperature.
Determine the zero drift and sensitivity drift per °C change in ambient
temperature.
Solution At 20°C, deflection/load characteristic is a straight line.
Sensitivity = 20 mm/kg.
At 30°C, deflection/load characteristic is still a straight line.
Sensitivity = 22 mm/kg.
Bias (zero drift) = 5 mm (the no-load deflection)
Sensitivity drift = 2 mm/kg
Zero drift/°C = 5/10 = 0.5 mm/°C
Sensitivity drift/°C = 2/10 = 0.2 (mm per kg)/°C
No comments:
Post a Comment
Tell your requirements and How this blog helped you.