14.2.1 Thermocouple tables
Although the preceding discussion has
suggested that the unknown temperature T can be evaluated from the calculated
value of the e.m.f. E1 at the hot junction using equation (14.1),
this is very difficult to do in practice because equation (14.1) is a high
order polynomial expression. An approximate translation between the value of E1
and temperature can be achieved by expressing equation (14.1) in graphical form
as in Figure 14.1. However, this is not usually of sufficient accuracy, and it
is normal practice to use tables of e.m.f. and temperature values known as
thermocouple tables. These include compensation for the effect of the e.m.f.
generated at the reference junction (Eref), which is assumed to be
at 0°C. Thus, the tables are only valid when the reference junction is exactly
at this temperature. Compensation for the case where the reference junction
temperature is not at zero is considered later in this section.
Tables for a range of standard
thermocouples are given in Appendix 4. In these tables, a range of temperatures
is given in the left-hand column and the e.m.f. output for each standard type
of thermocouple is given in the columns to the right. In practice, any general
e.m.f. output measurement taken at random will not be found exactly in the
tables, and interpolation will be necessary between the values shown in the table.
Example 14.1
If the e.m.f. output measured from a
chromel–constantan thermocouple is 13.419 mV with the reference junction at
0°C, the appropriate column in the tables shows that this corresponds to a hot
junction temperature of 200°C.
Example 14.2
If the measured output e.m.f. for a
chromel–constantan thermocouple (reference junc[1]tion
at 0°C) was 10.65 mV, it is necessary to carry out linear interpolation between
the temperature of 160°C corresponding to an e.m.f. of 10.501 mV shown in the
tables and the temperature of 170°C corresponding to an e.m.f. of 11.222 mV.
This interpolation procedure gives an indicated hot junction temperature of
162°C.
14.2.2 Non-zero reference junction
temperature
If the reference junction is immersed
in an ice bath to maintain it at a temperature of 0°C so that thermocouple
tables can be applied directly, the ice in the bath must be in a state of just
melting. This is the only state in which ice is exactly at 0°C, and otherwise
it will be either colder or hotter than this temperature. Thus, maintaining the
reference junction at 0°C is not a straightforward matter, particularly if the
environ[1]mental
temperature around the measurement system is relatively hot. In consequence, it
is common practice in many practical applications of thermocouples to maintain
the reference junction at a non-zero temperature by putting it into a
controlled environment maintained by an electrical heating element. In order to
still be able to apply thermocouple tables, correction then has to be made for
this non-zero reference junction temperature using a second thermoelectric law
known as the law of intermediate temperatures. This states that:
E(Th,T0)
= E(Th,Tr) + E(Tr,T0) (14.6)
where: E(Th,T0)
is the e.m.f. with the junctions at temperatures Th and T0,
E(Th,Tr) is the e.m.f. with the junctions at temperatures
Th and Tr, and E(Tr,T0)is the
e.m.f. with the junctions at temperatures Tr and T0, Th
is the hot junction measured temperature, T0 is 0°C and Tr
is the non-zero reference junction temperature that is somewhere between T0
and Th.
Example 14.3
Suppose that the reference junction
of a chromel–constantan thermocouple is main[1]tained
at a temperature of 80°C and the output e.m.f. measured is 40.102 mV when the
hot junction is immersed in a fluid.
The quantities given are Tr = 80°C
and E(Th,Tr) = 40.102 mV
From the tables, E(Tr,T0)
= 4.983 mV
Now applying equation (14.6), E(Th,T0)
= 40.102 + 4.983 = 45.085 mV
Again referring to the tables, this
indicates a fluid temperature of 600°C.
In using thermocouples, it is
essential that they are connected correctly. Large errors can result if they
are connected incorrectly, for example by interchanging the extension leads or
by using incorrect extension leads. Such mistakes are particularly serious
because they do not prevent some sort of output being obtained, which may look
sensible even though it is incorrect, and so the mistake may go unnoticed for a
long period of time. The following examples illustrate the sort of errors that
may arise:
Example 14.4
This example is an exercise in the
use of thermocouple tables, but it also serves to illustrate the large errors
that can arise if thermocouples are used incorrectly. In a particular
industrial situation, a chromel–alumel thermocouple with chromel–alumel extension
wires is used to measure the temperature of a fluid. In connecting up this
measurement system, the instrumentation engineer responsible has inadvertently
inter[1]changed the
extension wires from the thermocouple. The ends of the extension wires are held
at a reference temperature of 0°C and the output e.m.f. measured is 14.1 mV. If
the junction between the thermocouple and extension wires is at a temperature
of 40°C, what temperature of fluid is indicated and what is the true fluid
temperature?
Solution The initial step necessary
in solving a problem of this type is to draw a diagrammat[1]ical
representation of the system and to mark on this the e.m.f. sources,
temperatures etc., as shown in Figure 14.5. The first part of the problem is
solved very simply by looking up in thermocouple tables what temperature the
e.m.f. output of 12.1 mV indicates for a chromel–alumel thermocouple. This is
297.4°C. Then, summing e.m.f.s around the loop:
V = 12.1 = E1 + E2
+ E3 or E1 = 12.1 - E2 - E3
E2 = E3 =
e.m.f.(alumel-chromel)40 = -e.m.f.(chromel-alumel)40 * = -
1.611 mV
Hence:
E1 = 12.1 + 1.611
+ 1.611 = 15.322 mV
Interpolating from the thermocouple
tables, this indicates that the true fluid temperature is 374.5°C.
Example 14.5
This example also illustrates the
large errors that can arise if thermocouples are used incorrectly. An
iron–constantan thermocouple measuring the temperature of a fluid is connected
by mistake with copper–constantan extension leads (such that the two constantan
wires are connected together and the copper extension wire is connected to the
iron thermocouple wire). If the fluid temperature was actually 200°C, and the junction
between the thermocouple and extension wires was at 50°C, what e.m.f. would be
measured at the open ends of the extension wires if the reference junction is
main[1]tained at 0°C?
What fluid temperature would be deduced from this (assuming that the connection
mistake was not known about)?
Solution
Again, the initial step necessary is
to draw a diagram showing the junctions, tempera[1]tures
and e.m.f.s, as shown in Figure 14.6. The various quantities can then be calculated:
No comments:
Post a Comment
Tell your requirements and How this blog helped you.