6.2.3 Electrodynamic meters
Electrodynamic meters (or
dynamometers) can measure both d.c. signals and a.c. signals up to a frequency
of 2 kHz. As illustrated in Figure 6.4, the instrument has a moving circular
coil that is mounted in the magnetic field produced by two sepa[1]rately wound,
series-connected, circular stator coils. The torque is dependent upon the
mutual inductance between the coils and is given by:
T =
I1I2 dM/d0 (6.5)
where I1 and I2
are the currents flowing in the fixed and moving coils, M is the mutual
inductance and 0 represents the angular
displacement between the coils.
When used as an ammeter, the measured current is applied to both coils. The torque is thus proportional to current2. If the measured current is a.c., the meter is unable to follow the alternating torque values and it displays instead the mean value of current2. By suitable drawing of the scale, the position of the pointer shows the squared root of this value, i.e. the r.m.s. current.
Electrodynamic meters are typically
expensive but have the advantage of being more accurate than moving-coil and
moving-iron instruments. Voltage, current and power can
all be measured if the fixed and
moving coils are connected appropriately. When used for voltage measurement,
the instrument can typically measure voltages in the range of 0 to 30 volts.
However, it can be modified to measure higher voltages by placing a resistance
in series with it, as in the case of moving-coil and moving-iron meters. Also,
as in the moving-iron meter, a series resistance is particularly beneficial in
a.c. signal measurements because it compensates for the effect of coil
inductance by reducing the total resistance/inductance ratio, and hence
measurement accuracy is improved. This series resistance can either be inside
or outside the instrument case, as discussed above for the case of moving-iron
meters.
6.2.4 Clamp-on meters
These are used for measuring circuit
currents and voltages in a non-invasive manner that avoids having to break the
circuit being measured. The meter clamps onto a current-carrying conductor, and
the output reading is obtained by transformer action. The principle of
operation is illustrated in Figure 6.5, where it can be seen that the clamp-on
jaws of the instrument act as a transformer core and the current-carrying
conductor acts as a primary winding. Current induced in the secondary winding
is rectified and applied to a moving-coil meter. Although it is a very
convenient instrument to use, the clamp-on meter has low sensitivity and the
minimum current measurable is usually about 1 amp.
No comments:
Post a Comment
Tell your requirements and How this blog helped you.