google.com, pub-4497197638514141, DIRECT, f08c47fec0942fa0 Industries Needs: 7 Variable conversion elements

Saturday, December 18, 2021

7 Variable conversion elements

 7.1.2 Deflection-type d.c. bridge

A deflection-type bridge with d.c. excitation is shown in Figure 7.2. This differs from the Wheatstone bridge mainly in that the variable resistance Rv is replaced by a fixed resistance R1 of the same value as the nominal value of the unknown resistance Ru. As the resistance Ru changes, so the output voltage V0 varies, and this relationship between V0 and Ru must be calculated.

This relationship is simplified if we again assume that a high impedance voltage measuring instrument is used and the current drawn by it, Im, can be approximated to zero. (The case when this assumption does not hold is covered later in this section.) The analysis is then exactly the same as for the preceding example of the Wheatstone bridge, except that Rv is replaced by R1. Thus, from equation (7.1), we have:

                    V0 = Vi(Ru/(Ru + R3) - R1/(R1 + R2)            (7.3)

When Ru is at its nominal value, i.e. for Ru = R1, it is clear that V0 = 0 (since R2 = R3). For other values of Ru, V0 has negative and positive values that vary in a non-linear way with Ru.

Example 7.1

A certain type of pressure transducer, designed to measure pressures in the range 0–10 bar, consists of a diaphragm with a strain gauge cemented to it to detect diaphragm


deflections. The strain gauge has a nominal resistance of 120 Ω and forms one arm of a Wheatstone bridge circuit, with the other three arms each having a resistance of 120 Ω. The bridge output is measured by an instrument whose input impedance can be assumed infinite. If, in order to limit heating effects, the maximum permissible gauge current is 30 mA, calculate the maximum permissible bridge excitation voltage. If the sensitivity of the strain gauge is 338 mΩ /bar and the maximum bridge excitation voltage is used, calculate the bridge output voltage when measuring a pressure of 10 bar.

Solution

This is the type of bridge circuit shown in Figure 7.2 in which the components have the following values:

                            R1 = R2 = R3 = 120 Ω

Defining I1 to be the current flowing in path ADC of the bridge, we can write:

                           Vi = I1/(Ru + R3)

At balance, Ru = 120 and the maximum value allowable for I1 is 0.03 A.

Hence:

                         Vi = 0.03(120 + 120) = 7.2 V

Thus, the maximum bridge excitation voltage allowable is 7.2 volts.

For a pressure of 10 bar applied, the resistance change is 3.38 Ω, i.e. Ru is then equal to 123.38 Ω.

Applying equation (7.3), we can write:

V0 = Vi{Ru/(Ru + R3) - R1/(R1 + R2)} = 7.2 {123.38/243.38 – 120/240} = 50 mV

Thus, if the maximum permissible bridge excitation voltage is used, the output voltage is 50 mV when a pressure of 10 bar is measured.

The non-linear relationship between output reading and measured quantity exhibited by equation (7.3) is inconvenient and does not conform with the normal requirement for a linear input–output relationship. The method of coping with this non-linearity varies according to the form of primary transducer involved in the measurement system.

One special case is where the change in the unknown resistance Ru is typically small compared with the nominal value of Ru. If we calculate the new voltage V’0 when the resistance Ru in equation (7.3) changes by an amount Ru, we have:

V’0 = Vi{(Ru + Ru)/(Ru + Ru + R3) - R1/(R1 + R2)}                         (7.4)

The change of voltage output is therefore given by:

            V0 = V’0 - V0 = Vi Ru/(Ru + Ru + R3)

If Ru << Ru, then the following linear relationship is obtained:

                           V0/ Ru = Vi/(Ru + R3)                                        (7.5)

This expression describes the measurement sensitivity of the bridge. Such an approximation to make the relationship linear is valid for transducers such as strain gauges where the typical changes of resistance with strain are very small compared with the nominal gauge resistance.

However, many instruments that are inherently linear themselves at least over a limited measurement range, such as resistance thermometers, exhibit large changes in output as the input quantity changes, and the approximation of equation (7.5) cannot be applied. In such cases, specific action must be taken to improve linearity in the relationship between the bridge output voltage and the measured quantity. One common solution to this problem is to make the values of the resistances R2 and R3 at least ten times those of R1 and Ru (nominal). The effect of this is best observed by looking at a numerical example.

Consider a platinum resistance thermometer with a range of 0° –50°C, whose resistance at 0°C is 500 Ω and whose resistance varies with temperature at the rate of 4 Ω/°C. Over this range of measurement, the output characteristic of the thermometer itself is nearly perfectly linear. (N.B. The subject of resistance thermometers is discussed further in Chapter 14.)

Taking first the case where R1 = R2 = R3 = 500 Ω and Vi = 10 V, and applying equation (7.3):

                At 0°C; V0 = 0

At 25°C;    Ru = 600 Ω and V0 = 10(600/1100 – 500/1000) = 0.455 V

At 50°C;    Ru = 700 Ω and V0 = 10(700/1200 – 500/1000) = 0.833 V

This relationship between V0 and Ru is plotted as curve (a) in Figure 7.3 and the non-linearity is apparent. Inspection of the manner in which the output voltage V0 above changes for equal steps of temperature change also clearly demonstrates the non-linearity.

                    For the temperature change from 0 to 25°C, the change in V0 is

                                   (0.455 – 0) = 0.455 V

                   For the temperature change from 25 to 50°C, the change in V0 is

                                   (0.833 - 0.455) = 0.378 V

If the relationship was linear, the change in V0 for the 25–50°C temperature step would also be 0.455 V, giving a value for V0 of 0.910 V at 50°C.

Now take the case where R1 = 500 Ω but R2 = R3 D 5000 Ω and let Vi = 26.1 V:

       At 0°C;     V0 = 0

       At 25°C;     Ru = 600 Ω and V0 = 26.1(600/5600 – 500/5500) = 0.424 V

       At 50°C;      Ru = 700 Ω and V0 = 26.1(700/5700 – 500/5500) = 0.833 V

This relationship is shown as curve (b) in Figure 7.3 and a considerable improvement in linearity is achieved. This is more apparent if the differences in values for V0 over the two temperature steps are inspected.

                              From 0 to 25°C,    the change in V0 is 0.424 V

                               From 25 to 50°C, the change in V0 is 0.409 V

The changes in V0 over the two temperature steps are much closer to being equal than before, demonstrating the improvement in linearity. However, in increasing the values of R2 and R3, it was also necessary to increase the excitation voltage from 10 V to 26.1 V to obtain the same output levels. In practical applications, Vi would normally be set at the maximum level consistent with the limitation of the effect of circuit heating in order to maximize the measurement sensitivity (V0/ Ru relationship). It would therefore not be possible to increase Vi further if R2 and R3 were increased, and the general effect of such an increase in R2 and R3 is thus a decrease in the sensitivity of the measurement system.

The importance of this inherent non-linearity in the bridge output relationship is greatly diminished if the primary transducer and bridge circuit are incorporated as elements within an intelligent instrument. In that case, digital computation is applied


to produce an output in terms of the measured quantity that automatically compensates for the non-linearity in the bridge circuit.

Case where current drawn by measuring instrument is not negligible

For various reasons, it is not always possible to meet the condition that the impedance of the instrument measuring the bridge output voltage is sufficiently large for the current drawn by it to be negligible. Wherever the measurement current is not negligible, an alternative relationship between the bridge input and output must be derived that takes the current drawn by the measuring instrument into account.

Thevenin’s theorem is again a useful tool for this purpose. Replacing the voltage ´ source Vi in Figure 7.4(a) by a zero internal resistance produces the circuit shown in Figure 7.4(b), or the equivalent representation shown in Figure 7.4(c). It is apparent from Figure 7.4(c) that the equivalent circuit resistance consists of a pair of parallel resistors Ru and R3 in series with the parallel resistor pair R1 and R2. Thus, RDB is


given by:

                         RDB = - R1R2/(R1 + R2) + RuR3/(Ru + R3)           (7.6)

The equivalent circuit derived via Thevenin’s theorem with the resistance ´ Rm of the measuring instrument connected across the output is shown in Figure 7.4(d). The open-circuit voltage across DB, E0, is the output voltage calculated earlier (equation 7.3) for the case of Rm = 0:

                  E0 = V(Ru/(Ru + R3) - R1/(R1 + R2)                      (7.7)

If the current flowing is Im when the measuring instrument of resistance Rm is connected across DB, then, by Ohm’s law, Im is given by:

                           Im = E0/(RDB + Rm)                                   (7.8)

If Vm is the voltage measured across Rm, then, again by Ohm’s law:

                         Vm = ImRm = E0Rm/(RDB + Rm)                    (7.9)

Substituting for E0 and RDB in equation (7.9), using the relationships developed in equations (7.6) and (7.7), we obtain:

Vm = Vi [Ru/(Ru + R3) - R1/(R1 + R2)] Rm /R1R2/(R1 + R2) + RuR3/(Ru + R3) + Rm

Simplifying:

Vm = ViRm(RuR2 - R1R3) / R1R2(Ru + R3) + RuR3(R1 + R2) + Rm(R1 + R2)(Ru + R3)   (7.10)

Example 7.2

A bridge circuit, as shown in Figure 7.5, is used to measure the value of the unknown resistance Ru of a strain gauge of nominal value 500 Ω. The output voltage measured across points DB in the bridge is measured by a voltmeter. Calculate the measurement sensitivity in volts/ohm change in Ru if

(a) the resistance Rm of the measuring instrument is neglected, and

(b) account is taken of the value of Rm

Solution

For Ru = 500 , Vm = 0.

To determine sensitivity, calculate Vm for Ru = 501 Ω.

(a) Applying equation (7.3):    Vm = Vi [Ru/(Ru + R3) - R1/(R1 + R2)]


Substituting in values: Vm = 10 (501/1001 – 500/1000) = 5.00 mV

Thus, if the resistance of the measuring circuit is neglected, the measurement sensitivity is 5.00 mV per ohm change in Ru.

(b) Applying equation (7.10) and substituting in values:

Vm = 10 × 104 × 500(501 – 500) / 5002 (1001) + 500 × 501(1000) + 104 × 1000 × 1001 = 4.76 mV

Thus, if proper account is taken of the 10 kΩ value of the resistance of Rm, the true measurement sensitivity is shown to be 4.76 mV per ohm change in Ru.



No comments:

Post a Comment

Tell your requirements and How this blog helped you.

Labels

ACTUATORS (10) AIR CONTROL/MEASUREMENT (38) ALARMS (20) ALIGNMENT SYSTEMS (2) Ammeters (12) ANALYSERS/ANALYSIS SYSTEMS (33) ANGLE MEASUREMENT/EQUIPMENT (5) APPARATUS (6) Articles (3) AUDIO MEASUREMENT/EQUIPMENT (1) BALANCES (4) BALANCING MACHINES/SERVICES (1) BOILER CONTROLS/ACCESSORIES (5) BRIDGES (7) CABLES/CABLE MEASUREMENT (14) CALIBRATORS/CALIBRATION EQUIPMENT (19) CALIPERS (3) CARBON ANALYSERS/MONITORS (5) CHECKING EQUIPMENT/ACCESSORIES (8) CHLORINE ANALYSERS/MONITORS/EQUIPMENT (1) CIRCUIT TESTERS CIRCUITS (2) CLOCKS (1) CNC EQUIPMENT (1) COIL TESTERS EQUIPMENT (4) COMMUNICATION EQUIPMENT/TESTERS (1) COMPARATORS (1) COMPASSES (1) COMPONENTS/COMPONENT TESTERS (5) COMPRESSORS/COMPRESSOR ACCESSORIES (2) Computers (1) CONDUCTIVITY MEASUREMENT/CONTROL (3) CONTROLLERS/CONTROL SYTEMS (35) CONVERTERS (2) COUNTERS (4) CURRENT MEASURMENT/CONTROL (2) Data Acquisition Addon Cards (4) DATA ACQUISITION SOFTWARE (5) DATA ACQUISITION SYSTEMS (22) DATA ANALYSIS/DATA HANDLING EQUIPMENT (1) DC CURRENT SYSTEMS (2) DETECTORS/DETECTION SYSTEMS (3) DEVICES (1) DEW MEASURMENT/MONITORING (1) DISPLACEMENT (2) DRIVES (2) ELECTRICAL/ELECTRONIC MEASUREMENT (3) ENCODERS (1) ENERGY ANALYSIS/MEASUREMENT (1) EQUIPMENT (6) FLAME MONITORING/CONTROL (5) FLIGHT DATA ACQUISITION and ANALYSIS (1) FREQUENCY MEASUREMENT (1) GAS ANALYSIS/MEASURMENT (1) GAUGES/GAUGING EQUIPMENT (15) GLASS EQUIPMENT/TESTING (2) Global Instruments (1) Latest News (35) METERS (1) SOFTWARE DATA ACQUISITION (2) Supervisory Control - Data Acquisition (1)