7.7 Phase measurement
Instruments that convert the measured
variable into a phase change in a sinusoidal electrical signal include the
transit-time ultrasonic flowmeter, the radar level sensor, the LVDT and the
resolver. The most accurate instrument for measuring the phase difference
between two signals is the electronic counter-timer. However, two other methods
also exist that are less accurate but are nevertheless very useful in some
circumstances. One method involves plotting the signals on an X–Y plotter and
the other uses a dual beam oscilloscope.
7.7.1 Electronic counter-timer
In principle, the phase difference
between two sinusoidal signals can be determined by measuring the time that
elapses between the two signals crossing the time axis. However, in practice,
this is inaccurate because the zero crossings are susceptible to noise
contamination. The normal solution to this problem is to amplify/attenuate the
two signals so that they have the same amplitude and then measure the time that
elapses between the two signals crossing some non-zero threshold value.
The basis of this method of phase
measurement is a digital counter-timer with a quartz-controlled oscillator
providing a frequency standard that is typically 10 MHz. The crossing points of
the two signals through the reference threshold voltage level are applied to a
gate that starts and then stops pulses from the oscillator into an electronic
counter, as shown in Figure 7.20. The elapsed time, and hence phase difference,
between the two input signals is then measured in terms of the counter display.
7.7.2 X–Y plotter
This is a useful technique for
approximate phase measurement but is limited to low frequencies because of the
very limited bandwidth of an X–Y plotter. If two input
signals of equal magnitude are
applied to the X and Y inputs of a plotter, the plot obtained is an ellipse, as
shown in Figure 7.21. If the X and Y inputs are given by:
VX = V sin(ωt); VY = V sin(ωt +
At t = 0, VX = 0 and VY
= V sin
sin
Solution of equation (7.4) gives four
possible values for but the ambiguity
about which quadrant
7.7.3 Oscilloscope
Approximate measurement of the phase
difference between signals can be made using a dual-beam oscilloscope. The two
signals are applied to the two oscilloscope inputs and a suitable timebase
chosen such that the time between the crossing points of the two signals can be
measured. The phase difference of both low- and high-frequency signals can be
measured by this method, the upper frequency limit measurable being dictated by
the bandwidth of the oscilloscope (which is normally very high).
7.7.4 Phase-sensitive detector
The phase-sensitive detector
described earlier in section 5.5.9 can be used to measure the phase difference
between two signals that have an identical frequency. This can be exploited in
measurement devices like the varying-phase output resolver (see Chapter 20).
No comments:
Post a Comment
Tell your requirements and How this blog helped you.