17.5 Ultrasonic level gauge
Ultrasonic level measurement is one
of a number of non-contact techniques available. The principle of the
ultrasonic level gauge is that energy from an ultrasonic source above the
liquid is reflected back from the liquid surface into an ultrasonic energy
detector, as illustrated in Figure 17.4. Measurement of the time of flight
allows the liquid level to be inferred. In alternative versions, the ultrasonic
source is placed at the bottom of the vessel containing the liquid, and the
time of flight between emission, reflection off the liquid surface and
detection back at the bottom of the vessel is measured.
Ultrasonic techniques are especially
useful in measuring the position of the inter[1]face
between two immiscible liquids contained in the same vessel, or measuring the
sludge or precipitate level at the bottom of a liquid-filled tank. In either
case, the method employed is to fix the ultrasonic transmitter–receiver
transducer at a known height in the upper liquid, as shown in Figure 17.5. This
establishes the level of the liquid/liquid or liquid/sludge level in absolute
terms. When using ultrasonic instruments, it is essential that proper
compensation is made for the working temperature if this differs from the
calibration temperature, since the speed of ultrasound through air varies with
temperature (see Chapter 13). Ultrasound speed also has a small sensitivity to
humidity, air pressure and carbon dioxide concentration, but these factors are
usually insignificant. Temperature compensation can be achieved in two ways.
Firstly, the operating temperature can be measured and an appropriate
correction made. Secondly, and preferably, a comparison method can be used in
which the system is calibrated each time it is used by measuring the transit
time of ultrasonic energy between two known reference points. This second
method takes account of humidity, pressure and carbon dioxide concentration
variations as well as providing temperature compensation. With appropriate
care, measurement uncertainty can be reduced to about ±1%.
17.6 Radar (microwave) methods
Level-measuring instruments using
microwave radar are an alternative technique for non-contact measurement.
Currently, they are still very expensive (approximately £3000), but prices are
falling and usage is expanding rapidly. They are able to provide successful
level measurement in applications that are otherwise very difficult, such as
measurement in closed tanks, measurement where the liquid is turbulent, and
measurement in the presence of obstructions and steam condensate. The technique
involves directing a constant-amplitude, frequency-modulated microwave signal
at the liquid surface. A receiver measures the phase difference between the
reflected signal and the original signal transmitted directly through air to
it, as shown in Figure 17.6. This measured phase difference is linearly
proportional to the liquid level. The system is similar in principle to
ultrasonic level measurement, but has the important advantage that the
transmission time of radar through air is almost totally unaffected by ambient
temperature and pressure fluctuations. However, as the microwave frequency is
within the band used for radio communications, strict conditions on amplitude
levels have to be satisfied, and the appropriate licences have to be obtained.
17.7 Radiation methods
The radiation method is an expensive
technique, which uses a radiation source and detector system located outside a
liquid-filled tank in the manner shown in Figure 17.6. The non-invasive nature
of this technique in using a source and detector system outside
the tank is particularly attractive.
The absorption of both beta rays and gamma rays varies with the amount of
liquid between the source and detector, and hence is a function of liquid
level. Caesium-137 is a commonly used gamma-ray source. The radiation level
measured by the detector I is related to the length of liquid in the path x
according to:
where I0 is the intensity of
radiation that would be received by the detector in the absence of any liquid, ยต is the mass absorption coefficient for the liquid and p is the mass
density of the liquid.
In the arrangement shown in Figure
17.7, the radiation follows a diagonal path across the liquid, and therefore
some trigonometrical manipulation has to be carried out to determine the liquid
level h from x. In some applications, the radiation source can be located in the
centre of the bottom of the tank, with the detector vertically above it. Where
this is possible, the relationship between the radiation detected and liquid
level is obtained by directly substituting h in place of x in equation (17.2).
Apart from use with liquids at normal temperatures, this method is commonly
used for measuring the level of hot, liquid metals. However, because of the
obvious dangers associated with using radiation sources, very strict safety
regulations have to be satisfied when applying this technique. Very low
activity radiation sources are used in some systems to overcome safety problems
but the system is then sensitive to background radiation and special
precautions have to be taken regarding the provision of adequate shielding. Because
of the many difficulties in using this technique, it is only used in special
applications
No comments:
Post a Comment
Tell your requirements and How this blog helped you.