4.2 Control of calibration environment
Any instrument that is used as a standard in calibration procedures must
be kept solely for calibration duties and must never be used for other
purposes. Most particularly, it must not be regarded as a spare instrument that
can be used for process measurements if the instrument normally used for that
purpose breaks down. Proper provision for process instrument failures must be
made by keeping a spare set of process instruments. Standard calibration
instruments must be totally separate.
To ensure that these conditions are met, the calibration function must be
managed and executed in a professional manner. This will normally mean setting
aside a partic[1]ular place within
the instrumentation department of a company where all calibration operations
take place and where all instruments used for calibration are kept. As far as
possible this should take the form of a separate room, rather than a
sectioned-off area in a room used for other purposes as well. This will enable
better environmental control to be applied in the calibration area and will
also offer better protection against unauthorized handling or use of the
calibration instruments. The level of environmental control required during
calibration should be considered carefully with due regard to what level of
accuracy is required in the calibration procedure, but should not be
overspecified as this will lead to unnecessary expense. Full air conditioning
is not normally required for calibration at this level, as it is very
expensive, but sensible precautions should be taken to guard the area from
extremes of heat or cold, and also good standards of cleanliness should be
maintained. Useful guidance on the operation of standards facilities can be
found elsewhere (British Standards Society, 1979).
Whilst it is desirable that all calibration functions are performed in
this carefully controlled environment, it is not always practical to achieve
this. Sometimes, it is not convenient or possible to remove instruments from
process plant, and in these cases, it is standard practice to calibrate them in
situ. In these circumstances, appropriate corrections must be made for the
deviation in the calibration environmental conditions away from those
specified. This practice does not obviate the need to protect calibration
instruments and maintain them in constant conditions in a calibration
laboratory at all times other than when they are involved in such calibration
duties on plant.
As far as management of calibration procedures is concerned, it is
important that the performance of all calibration operations is assigned as the
clear responsibility of just one person. That person should have total control
over the calibration function, and be able to limit access to the calibration
laboratory to designated, approved personnel only. Only by giving this
appointed person total control over the calibration function can the function
be expected to operate efficiently and effectively. Lack of such definite
management can only lead to unintentional neglect of the calibration system,
resulting in the use of equipment in an out-of-date state of calibration and
subsequent loss of traceability to reference standards. Professional management
is essential so that the customer can be assured that an efficient calibration
system is in operation and that the accuracy of measurements is guaranteed.
Calibration procedures that relate in any way to measurements that are
used for quality control functions are controlled by the international standard
ISO 9000 (this subsumes the old British quality standard BS 5750). One of the
clauses in ISO 9000 requires that all persons using calibration equipment be
adequately trained. The manager in charge of the calibration function is
clearly responsible for ensuring that this condition is met. Training must be
adequate and targeted at the particular needs of the calibration systems
involved. People must understand what they need to know and especially why they
must have this information. Successful completion of training courses should be
marked by the award of qualification certificates. These attest to the
proficiency of personnel involved in calibration duties and are a convenient
way of demonstrating that the ISO 9000 training requirement has been satisfied.
No comments:
Post a Comment
Tell your requirements and How this blog helped you.